Hanna HI 93703 Portable
Turbidity Meter |
|
 |
|
| |
|

The HI 93703 portable
microprocessor-based turbidity meter provides lab-grade accuracy even in
the field. With 4 keys and weighing only 350 grams, HI 93703 measures
turbidity from 0 to 1000 FTU (NTU)*. For best field accuracy, it
measures from 0 to 50 FTU in steps of 1/100th of FTU. This is 10 times
better than comparable models on the market that measure 10 to 50 FTU
only in decimal points! The meter meets the ISO 7027 requirements.
HI 93703 is equipped with an infrared LED as
source of light. Unlike standard lamps, the infrared LED has a long
life. More significantly, it maintains constant emission for the entire
life of the instrument. The wavelength peaks at 890 nm, which provides
the required intensity of diffused light even in samples with low
turbidity values, and also reduces the interference from any colors. The
meter is very easy to use. All operations can be carried out with only
four keys and troubleshooting functions can be performed with error code
on LCD. Moreover, the meter's versatility and durability ensure low
maintenance. Sensitive to the user's need to work with properly
calibrated meters, Hanna Instruments has equipped HI 93703 with a unique
GLP (Good Laboratory Practice) feature. GLP circuitry automatically
stores and retrieves the last calibration data. At the touch of a key,
the last calibration data together with time and date are displayed
allowing the user to maintain an accurate calibration schedule Auto
shut-off turns the meter off after 5 minutes of non-use to save
batteries.
In addition to the outstanding features listed above, the
upgraded HI 93703-11 can memorize up to 200 measurements at any time.
These can be later recalled on the LCD or downloaded to a PC. HI 93703-11
has a real time clock and can be also calibrated at 500 FTU. Use the
optional Windows® compatible HI 92000 software to get the most from
your instrumentation. |
|
* 1 FTU (Formazine Turbidity Unit) = 1 NTU (Nephelometric Turbidity
Unit)
|
|
|
PRODUCT
SPECIFICATIONS

| Range |
0.00 to 50.00 FTU / 50
to 1000 FTU* |
| Resolution |
0.01 / 1 FTU |
| Accuracy |
±0.5 FTU or ±5%
(whichever greater) |
| Typical EMC
Deviation |
±2% Full Scale |
| Light
Source/Life |
High Emission Infrared
LED / Life of the instrument |
| Light Detector |
Silicon Photocell |
| Computer
Interface |
RS232, with optional HI
92000 software (HI 93703-11 only) |
| Battery
Type/Life |
4 x 1.5V AA batteries 60
hours or 900 measurements Auto-shut off after 5 minutes |
| Environment |
32 to 122°F (0 to 50°C);
RH 95% |
| Dimensions |
8.7 x 3.2 x 2.6" (220 x
82 x 66 mm) |
| Weight |
18 oz. (510 g) |
|
| |
|
PURCHASE TURBIDITY METERS
 |
|
ITEM NUMBER |
DESCRIPTION |
PRICE |
|
|
Hanna
Turbidity Meter with Maintenance Kit (HI 731313) |
|
|
|
Hanna
Turbidity Meter with RS232 |
|
|
|
Hanna
Turbidity Meter |
|
|
|
Replacement
Cuvets, 4 pack |
|
|
|
Replacement
Cuvet Caps, 4 pack |
|
|
|
Rugged
carrying case with calibration solutions HI 93703-0 and HI 93703-10, HI
93703-50 cleaning solution, HI 731318 tissue for wiping cuvets, 2
measurement cuvets |
|
|
|
AMCO-EPA-1
cal. sol. @ 0 FTU (30mL) |
|
|
|
AMCO-EPA-1
cal. sol. @ 500 FTU (30mL) for HI 93703-11 only |
|
|
|
AMCO-EPA-1
cal. sol. @ 10 FTU (30mL) |
|
|
|
Cleaning
solution for cuvets (230mL) |
|
|
|
Tissue for
wiping cuvets (4 pcs) |
|
|
|
|