LaMotte 2020 Turbidity Meter
  • Ideal choice for regulatory monitoring in the lab or field
  • Simple calibration procedure
  • RS-232 port allows use to interface with a datalogger or computer
  • Approx 450 battery life with low battery indicator
  • Tips
  • Troubleshooting
  • Maintenance
  • Manual
 

2020 TURBIDITY METER OVERVIEW

The 2020 sets a new standard for laboratory accuracy in an extremely compact, portable instrument for turbidity measurement. A multi-detector optical configuration assures long term stability and minimizes stray light and color interferences. All readings are determined by the process of signal averaging over a 5 second period. This minimizes fluctuations in readings attributed to large particles and results in rapid, highly repeatable measurements. The 2020 has a 1 year warranty and includes AC power adapter, 9V battery, 4 sample vials with screw caps, both 1.0 and 10.0 standards and operators manual - all in a sturdy case.

PURCHASE TURBIDITY METERS

ITEM NUMBER DESCRIPTION PRICE

QTY LLI1799

Turbidimeter 2020, Portable, 120 VAC
$0.00
REPLACEMENT ITEMS

QTY LLI1476

1.0 NTU AMCO Standard
$48.50

QTY LLI1477

10.0 NTU AMCO Standard
$78.00

QTY LLI1726110

RS-232 Cable (110v)
$24.55

QTY LLI02864

Cuvettes, 4 pack (15 mL)
$27.40

TURBIDITY METER SPECIFICATIONS

Range: 0 - 1100 NTU
Accuracy: ±2% of reading below 100 NTU,
±3% of reading above 100 NTU
Resolution: 0.01 NTU (0-9.99 NTU)
0.1 NTU (10-99.9 NTU)
1 NTU (100-1000 NTU)
Response Time: Less than 5 seconds
 

TIPS

  1. Samples should be collected in a clean glass or polyethylene container.

  2. Samples should be analyzed as soon as possible after collection.

  3. Discard tubes that are badly scratched.

  4. Gently mix sample by inverting before taking a reading but avoid introducing air bubbles.

  5. Turbidity readings will be affected by electric fields around motors.

  6. Carbon in the sample will absorb light and cause low readings.

  7. Observe shelf life recommendations for turbidity standards.

  8. The turbidimeter should be placed on a surface free from vibration. Vibrations can cause high readings.

  9. Excessive color in a sample will absorb light and cause high readings. The user should verify if a certain level of color will cause a significant error at the level of turbidity being tested.

TROUBLESHOOTING

Problem
Check
Action
No Power Battery Replace
AC Adapter Plug in
AC Wall Outlet Verify power source
Suspect Calibration Check calibration with standards Use new standards
Verify standards with Formazin Run alternate test with Formazin
Verify with another meter Check other meter calibrations
Check tube alignment Re-align tube
Check sample tubes for dirt and scratches Check, clean and/or replace if necessary
Check to see if internal meter components are wet Always dry tubes before inserting. Examine chamber for visible moisture.
Reset meter to factory calibration With meter off, hold down key and press READ
Battery A fresh battery should be installed at all times, even when using the power adapter. This will ensure that the meter will power down properly and the calibration will not be lost.
Er1 Very low battery Change battery
Er2 Over range Dilute sample
Er3 Burnt out bulb or misaligned tube Check alignment
BAT Low Battery Change battery
Notes:
  1. If the battery voltage is very low, the meter may read Er3 rather than Er1. Replace battery with new Duracell 9V battery.

  2. If unable to calibrate meter to standard solutions, try resetting the meter to factory calibration. To reset the calibration to factory settings: with the meter off, press and hold the DOWN button and push READ. The meter will turn on and the calibration will be reset.

MAINTENANCE

Replacing the Battery

The LaMotte 2020 uses a standard 9-volt alkaline battery that is available worldwide. A fresh battery should be installed at all times, even when using the power adapter. This will ensure that the meter will power down properly. The battery compartment is located on the bottom of the case. To replace the battery:

  1. Open the battery compartment lid

  2. Remove the battery and disconnect the battery from the polarized plug.

  3. Carefully connect the new battery to the polarized plug and insert it into the compartment.

  4. Close the battery compartment lid

Replacing the Lamp

The tungsten lamp included with the model 2020 has a life of approximately 800 hours. If the display becomes unstable when using LaMotte AMCOTM standards, call LaMotte Company for a return authorization number to have the lamp replaced and have the unit examined.