LaMotte 2020e Turbidity
Meter |
|
 |
- Patent-pending
optical design features focusing optics for low level precision and
accuracy
- MSP430
Microcontroller used in the most advanced controller on the market
for hand held applications
- Advanced
calibration algorithms
- Tube positioning
ring limits vial variability
- Easy menu driven
operation
- Large LCD
display
- 4000 point data
log
- Stored results
can be viewed directly or downloaded to a computer via RS232 cable
- Meets USEPA
design criteria as specified by USEPA method 180.1
- Kit is supplied
with 0, 1, and 10 NTU standard, sample bottle, 4 sample tubes, an
extra battery
- AC power adapter
is optional
|
|
|
|
 |

LLI1979 Manual |
|

Formazin Standard MSDS |
| |
|
2020e TURBIDITY METER OVERVIEW

The newly redesigned 2020e combines an advanced microprocessor with a
patent-pending optical chamber resulting in higher accuracy and a wider
range.
The multi-detector optical configuration
assures long term stability and minimizes stray light and color
interferences. All readings are determined by the process of
signal averaging over a 5 second period. This minimizes
fluctuations in readings attributed to large particles and results in
rapid highly repeatable measurements. Ideally suited for both
low-level drinking water applications as well as monitoring high
turbidity in the field. |
|
|
PURCHASE TURBIDITY METERS
 |
|
ITEM
NUMBER |
DESCRIPTION |
PRICE |
|
|
Turbidimeter
2020e Kit, Portable, 120 VAC |
|
REPLACEMENT ITEMS

|
|
|
0.0 NTU AMCO™
Standard |
|
|
|
1.0 NTU AMCO™
Standard |
|
|
|
10.0 NTU AMCO™
Standard |
|
|
|
RS-232 Cable
(110v) |
|
|
|
Power Adapter |
|
|
|
Vial Ring |
|
|
|
Replacement
Vials - 6 pack |
|
|
|
TURBIDITY METER SPECIFICATIONS
 |
| Turbidity
Specifications |
| Unit of Measure: |
NTU, FNU, FAU, ASBC, EBC |
| Range: |
0-4000 |
| Resolution: |
0.01 NTU/FNU 0.00-10.99, 0.1 NTU/FNU
11.00-109.9,
1 NTU/FNU 110-4000 |
| Accuracy: |
±2% 0-100 NTU, ±3% above 100 NTU |
| Detection Limit: |
0.05 NTU/FNU |
| Range Selection: |
Automatic |
| Reproducibility: |
0.02 NTU/FNU, 0.5 FAU |
| Light Source: |
Tungsten (2020e EPA) |
|
Meter Features:
|
| Signal Averaging: |
Disabled, 2, 5, 10 |
| Battery: |
9V |
| AC Power: |
Optional |
| Data Logging: |
4000 points |
| Auto Shut-Off: |
Disabled, 5, 10, 30 |
| Optional Software: |
SmartLink 2 |
| Languages: |
English, French, Spanish,
Japanese, Italian, Portuguese |
| Response Time: |
<5 Seconds |
|
|
|